A comparative study on the dosimetric characteristics of iViewGT a-Si EPIDs

  • Peter Winkler (Autor)
  • Alfred Hefner (Autor)
  • Georg Dietmar (Autor)

Aktivität: Vortrag ohne Tagungsband / VorlesungPräsentation auf einer wissenschaftlichen Konferenz / Workshop

Beschreibung

The use of an electronic portal imaging device (EPID) for dosimetric purposes requires knowledge of the detector´s dose-response behavior. The Elekta iViewGT amorphous silicon (a-Si) EPID generally exhibits a nonlinear response to radiation. Variations of EPID sensitivity with dose and dose rate, respectively, up to 10% are reported in literature 1,2. The aim of this study was to investigate the variation of the dosimetric properties of several iViewGT EPIDs and to deduce a minimum number of measurements, which are required for a complete dosimetric characterization of such devices.