MEASUREMENTS AND MONTE CARLO SIMULATIONS OF THE RESPONSE OF THE RADOS PERSONAL DOSEMETERS WITH MTS-N (LiF:Mg,Ti) AND MCP-N (LiF:Mg,Cu,P) THERMOLUMINESCENT DETECTORS TO X- AND GAMMA-RAYS

  • Barbara Obryk (Autor)
  • Theresa Dimovasili (Autor)
  • Marcin Budzanowski (Autor)
  • Pavel Bilski (Autor)
  • P. Olko (Autor)
  • Christian Hranitzky (Autor)
  • Hannes Stadtmann (Vortragender)

Aktivität: Vortrag ohne Tagungsband / VorlesungPräsentation auf einer wissenschaftlichen Konferenz / Workshop

Beschreibung

Introduction: The LADIS dosimetric service at the Institute of Nuclear Physics (IFJ) Kraków, LADIS, applies MCP-N (LiF:Mg,Cu,P) and MTS-N (LiF:Mg,Ti) thermoluminescence detectors placed in RADOS holder for individual monitoring of Hp(10) and Hp(0.07) in photon radiation fields. Three standard TL pellets (4.5 mm of diameter and 0.9 mm of thickness) are placed in the holder under the cover of 0.244 g/cm2 Al foil and the fourth detector, for Hp(0.07) remains uncovered.
Zeitraum8 Juli 200713 Juli 2007
Ereignistitel"Solid State Dosimetry 2007" (SSD15)
VeranstaltungstypSonstiges
BekanntheitsgradInternational

Research Field

  • Biosensor Technologies