Abstract
Electronic portal imaging devices (EPIDs) are a convenient tool for geometric as well as dosimetric IMRT verification. However, the use of such devices for dosimetric purposes requires a detailed knowledge of their dosimetric properties. The aim of this paper was to describe a method for acquiring portal dose images (PDIs) with an amorphous silicon (a-Si) EPID and to assess its accuracy. Three IMRT treatment plans in the energy range from 6 to 25 MV were verified on a field-by-field basis. Reference dose distributions were measured using radiographic films as well as a 2D matrix of ionization chambers.
| Originalsprache | Englisch |
|---|---|
| Titel | The 9th International Workshop on Electronic Portal Imaging |
| Seiten | 85-86 |
| Seitenumfang | 2 |
| Publikationsstatus | Veröffentlicht - 2006 |
| Veranstaltung | The 9th International Workshop on Electronic Portal Imaging - Dauer: 1 Jan. 2006 → … |
Konferenz
| Konferenz | The 9th International Workshop on Electronic Portal Imaging |
|---|---|
| Zeitraum | 1/01/06 → … |
Research Field
- Biosensor Technologies