Revisiting the concept of constant tissue conductivities for volume estimation in dialysis patients using bioimpedance spectroscopy

Daniel Schneditz, Sebastian Mussnig, Simon Krenn, Manfred Hecking

Publikation: Beitrag in FachzeitschriftArtikelBegutachtung

Abstract

Rationale: Current estimation of body fluid volumes in hemodialysis patients using bioimpedance analysis assumes constant specific electrical characteristics of biological tissues despite a large variation in plasma Na+ concentrations [Na+], ranging from 130 to 150 mmol/L. Here, we examined the potential effect of variable [Na+] on bioimpedance-derived volume overload. Method: Volumes were calculated from published whole-body extra- and intracellular resistance data and relationships using either "standard" or "revised" specific electrical characteristics modeled as functions of [Na+]. Result: With "standard" assumptions, volumes increased with increasing [Na+]. The increase in volume overload was about 0.5 dm(3) and 3% of extracellular volume per 10 mmol/dm(3) of [Na+] in a 75 kg patient. This increase was abolished when the same bioimpedance data were analyzed under "revised" conditions. Discussion: The overestimation in extracellular volume overload in the range of 0.5 dm(3) per 10 mmol/dm(3) [Na+] perfectly matches the positive relationship determined in a large cohort of hemodialysis patients. The bias may be considered moderate when interpreting data of individual patients, but may become important when comparing data of larger patient groups. The bias disappears when analysis of bioimpedance data accounts for differences in tissue electrical properties, using individual [Na+].
OriginalspracheEnglisch
Seiten (von - bis)67-73
Seitenumfang7
FachzeitschriftInternational Journal of Artificial Organs
Volume46
Issue2
Frühes Online-DatumDez. 2022
DOIs
PublikationsstatusVeröffentlicht - Feb. 2023

Research Field

  • Medical Signal Analysis

Fingerprint

Untersuchen Sie die Forschungsthemen von „Revisiting the concept of constant tissue conductivities for volume estimation in dialysis patients using bioimpedance spectroscopy“. Zusammen bilden sie einen einzigartigen Fingerprint.

Diese Publikation zitieren